18.1.3
Connection of analog output
-
Current output
Screen
Load resistance
510 Ω or less
ZERO SPAN
A-OUT
F.G.
A
-
Voltage output
Screen
Load resistance
5 kΩ or less
ZERO SPAN
A-OUT
F.G.
V
Note:
-
Use shielded cable and connect the shield to the F.G. terminal.
-
To conform with CE mark or JIS standards, use shielded cable and connect the shield
to the F.G. terminal.
18.1.4
Analog output target
The analog output target is selected by C function CF-70.
The analog output target can be selected from [Display interlock], [Gross value] and [Net
value].
The default is [Display interlock].
18.1.5
Scaling of analog output
The analog output is set as the minimum value to the maximum value from 0 to 10 000
by setting CF-71 and CF-72.
Weight indicator CSD-903
18 Options
EN-166
Minebea Intec