Set up an operation control panel if the instrument is used in the following locations:
-
Places exposed to water and/or spillages
-
Places contaminated with dust and/or ine particles
NOTICE
Risk of malfunctions
Use the instrument under the following conditions of temperature and humidity.
Environmental temperature: -10°C to 50°C (Preservation -20°C to 60°C). The
temperature range of JIS B 7611-2 conformity is 10 °C to 40 °C.
Environmental humidity: Less than 85 % R. H. or less (Non condensing)
2.3.2
Installing the instrument
NOTICE
Warning of damage to property and/or the environment.
Install CSD-903 based on the following dimensions.
Ensure suicient space around the instrument.
The following are the dimensions of CSD-903 and the clearance required.
Front
144
72
Rear dimension
137
67
Side
10.2
136.5
14
Panel cut
136
68
+1
0
+
1
0
all dimensions in mm
The following are the side dimensions when BCD output (2) is connected.
Other connectors attached for external control input/output (1) have the same
dimensions as the BCD output.
Weight indicator CSD-903
2 Safety instructions
EN-12
Minebea Intec
Содержание CSD-903