①
Change to digital linearization mode
②
Compensation at point
②
Compensation at point 1
②
Compensation at point 2
⑤
Compensation at point 2
⑥
Stability check at point 3
⑤
Compensation at point 3
⑤
Quit digital linearization mode
Note:
-
Calibrating by transmission
(-RS-) is displayed.
-
When the calibration is force inished, all data returns to its condition prior to
calibration without registration.
-
Be sure to register the data for calibration with a weight after conirming the
condition is stable.
17.1.3.1
Switch to digital linearization mode
Switch to digital linearization mode by writing [11] in holding register address 56833.
17.1.3.2
Stability check at point 1
Place weights not exceeding the maximum load on the load cell. Stable/Unstable is read
from holding register address 56845.
0: Unstable
1: Stable
17.1.3.3
Compensation at point 1
The weight at point 1 is set and registered in holding register address 56851 to 56852.
Setting range: 1 to maximum display value
LN-L error
[24] (LN-L) is written in input register address 52 when input load value is lower than the
zero point.
Weight indicator CSD-903
17 ModBus communication
EN-160
Minebea Intec
Содержание CSD-903
Страница 243: ......