AccuPyc II 1340
System Commands
Aug 2013
3-37
Zero
This command enables you to zero the pressure transducer or the volume offset.
The pressure transducer zeroes automatically before each cycle in an analysis or calibration; therefore,
it is not necessary for you to zero for these operations. Typically, this function is not required unless
you manually perform analyses for an extended period of time.
Press
Alt + 0
to access the Zero function.
Enables you to zero the pressure transducer or the volume
offset.
Choices
:
Pressure, Volume
If you choose
Volume
, be sure the sample chamber is empty.
Press
CHOICE
until the desired option is displayed, then
press
ENTER
.
If you choose
Pressure
, the next prompt is not displayed.
Does not display for Pressure
Displays when you choose
Volume
and
if you selected
Yes
at the
Request insert
prompt in Report Options.
Choices:
None, 0.1 cm
3
(for 1 cm
3
pycnometer)
None, 3.5 cm
3
, 1.0 cm
3
(for 10 cm
3
pycnometer)
None, 35 cm
3
, 10 cm
3
(for 100 cm
3
pycnometer)
None, 650 cm
3
, 1300 cm
3
(for 2000 cm
3
pycnometer)
Press
CHOICE
until the appropriate insert is displayed, then
press
ENTER
.
Press
ENTER
to begin the zero operation or
Alt + CLEAR
to
cancel the operation.
If you are zeroing the
Volume
, the purges and cycles
specified in the current analysis parameters are used to
calculate the volume offset.
After the operation is complete, the pycnometer returns to
the
Reload
prompt.
Unit[n]
SN1234 10 cm3
X - X - X
Zero Type?
Pressure
Unit[n]
SN1234 10 cm3
Zero Volume
Chamber insert?
None
Unit[n]
SN1234 10 cm3
Zero Volume
[Enter] to start
[Escape] to cancel
or
Unit[n]
SN1234 10 cm3
Zero Pressure
[Enter] to start
[Escape] to cancel
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...