System Commands
AccuPyc II 1340
3-14
Aug 2013
Report Options
The Report options command enables you to specify the report mode (density or volume for analysis)
and the report destination.
This prompt displays when you selected
Yes
for run precision
and if you have requested more than five cycles.
Enter the run precision volume tolerance which is expressed
as a percent of nominal cell volume (sample capacity), then
press
ENTER
.
Range:
0.01% to 50%
Press
CHOICE
until
Report Options
is displayed, then press
ENTER
.
Press
CHOICE
to select
Density
or
Volume
, the mode in
which the analysis report will be displayed on the
pycnometer.
If you select
Density
, a
Sample Mass?
prompt displays
during the Analysis function so that you can enter a value for
the sample mass. For accurate data, the sample mass must be
entered when choosing
Density
.
Press
ENTER
.
A sample ID (identification) is a unique identifier of the
sample. Using sample IDs can help you keep track of data
from multiple analyses. For example, you can use the
sample ID as a date and time stamp. A
Sample ID
prompt
displays during the analysis operation, allowing you to enter
the identification when you select
Yes
to this prompt.
Choices:
Yes, No
Press
CHOICE
to select
Yes
or
No
, then press
ENTER
.
Unit[n]
SN1234 10 cm3
Analysis Parameters
Percent Full-Scale:
0.05%
Unit[n]
SN1234 10 cm3
Set-up Type?
Report Options
Unit[n]
SN1234 10 cm3
Report Options
Anls Display Mode?
Density
Unit[n]
SN1234 10 cm3
Report Options
Request Sample ID?
Yes
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...