Installing the Analyzer
AccuPyc II 1340
2-8
June 2013
Turning On and Off the Pycnometer
Place the power ON/OFF switch on the rear panel in the ON ( | ) position. After a few seconds,
the system vents automatically and the green indicator light on the front panel illuminates.
The following prompt is displayed in the keypad window:
Refer to
for a description of the pycnometer display.
Allow approximately 30 minutes for the analyzer to warm before performing analyses. For
analyses that require very precise results, allow the analyzer to warm a minimum of two hours.
It is recommended that you leave the analyzer on at all times.
If it does become necessary to turn it off, perform the following sequence of steps.
1.
Allow any analyses in progress to complete.
2.
Place the printer ON/OFF switch in the OFF position.
3.
Place the analyzer ON/OFF switch in the OFF position
.
Do not turn off the analyzer while initialization is in progress. Doing so may damage
the instrument.
Unit[n]
SN1234
10 cm3
X - X - X
Reload
Pressure = (current pressure)
It is important that a constant temperature be maintained inside the unit because a
change in temperature could alter analysis results. We recommend that the
pycnometer remain turned on at all times to maintain thermal stability. Power
consumption is small and the cost of electricity is minimal.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...