AccuPyc II 1340
Appendix A
Oct 2013
A-3
ANLSERR: Underrange
NN/ZZ cycles completed
Automatic operation
has been canceled
CAL_ERR: Reset Pressure Cal. ?
CAL_ERR: Temperature
Calibration Failed
Cause:
A pressure underrange occurred during an analysis because an error
occurred in the pressure measurement electronics.
Action:
Call a Micromeritics Service Representative.
Cause:
The automatic operation has been canceled by the user.
Action:
Wait for the termination process to complete or end the termination
process by pressing
Alt + CLEAR
two more times.
Cause:
The calibration information could not be read from the command or
analysis module.
Action:
Choose
Yes
and press
Enter
to reset the pressure and temperature
calibration to nominal values. Resetting to nominal may restore
normal readings but a proper calibration should be performed by a
service technician. See the section
Reset Pressure Calibration
in the
System Commands
section.
Action:
Choose
No
to avoid resetting the calibration. The instrument will be
unresponsive.
Cause:
The temperature calibration was unsuccessful.
Action:
Contact your service representative.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...