AccuPyc II 1340
Appendix A
Oct 2013
A-7
DTA_WRN: Cal std
10% of full-scale
DTA_WRN: Volume
10% of full-scale
HW_ERR: Calibration
could not be read
Cause A:
You entered a value for the calibration standard that was less than 10%
of the nominal cell chamber volume. Resulting data may be inaccurate.
Action A:
Be sure you enter the value recorded on the Calibration Standard case.
Cause B:
You used an inappropriate calibration standard for the nominal cell
chamber.
Action B:
Use a calibration standard of sufficient size (calibration standards are
available from Micromeritics). The calibration standard should occupy
at least 10% of the nominal cell chamber volume and the more nearly
filled the cell is, the better the calibration.
Cause:
There was not enough sample placed in the sample cup to give an
accurate reading.
Action:
Rerun the sample using enough material to occupy the maximum
amount of the nominal cell chamber volume.
Cause:
The calibration information could not be read from the command or
analysis module.
Action:
Close the application and cycle power on the instrument. Contact your
service representative if the problem persists.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...