Appendix A
AccuPyc II 1340
A-10
Oct 2013
SYS_ERR: Power Fail
NN/ZZ Cycles Completed
TRN_ERR: Timeout
failed to respond
Updating Unit [n]
failed: Press [CLEAR]
USR_ERR: Cal std
10% of full-scale
Cause:
A power failure occurred and when power resumed, the automatic
operation was canceled.
Action:
Restart the automatic operation if desired.
Cause:
The receiving device took longer than 10 seconds to acknowledge
receipt of data from the pycnometer.
Action:
Make sure the receiving device is properly connected to the
pycnometer RS-232 port and is turned on. Verify that the serial I/O
parameters controlling the receiving device correspond with the data
transmission parameters in the set up mode.
Cause:
One of the analysis modules has malfunctioned.
Action:
Press
CLEAR
and follow the instructions displayed in the keypad
window.
After disconnecting the faulty unit (as instructed), you may reconnect
the control module with the remaining analysis module(s) and continue
using your system.
Cause:
You tried to enter a calibration standard volume that is less than 10% of
the nominal full-scale volume.
Action:
Enter a volume that represents at least 10% of the nominal cell
chamber volume.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...