AccuPyc II 1340
Preparing for Analysis
June 2013
4-3
5.
Weigh the cup and sample and record the weight. Subtract the empty sample cup weight
from the sample cup plus sample weight to determine the sample weight. (If volume only
is to be measured, you may skip this step.)
6.
Remove the cell chamber cap, place the sample cup (with sample) into the cell chamber,
then replace the cap.
It is best not to lay the cap down while loading the sample. Immediately replace the cap
when the sample is loaded. This will prevent particles from accumulating on the greased
surface (refer to
system components).
7.
If you have a temperature-controlled AccuPyc, perform the following steps before
proceeding with
Starting an Analysis
in the next section.
a. Turn on the bath circulator and specify the temperature you wish to use for your
analysis. (Refer to the manufacturer’s manual for instructions on operating the bath
circulator.)
b. Allow the temperature to stabilize to the set temperature. Observe the temperature
reading on the fourth line of the display.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...