5. Remove the protective caps from the ends of the gas entrance tubing, then remove the
ferrules from the upper end of the tubing.
6. Remove the nut from the center port and slide it onto the end of the gas entrance tubing,
then replace the front and rear ferrules. Be sure the ferrules are oriented as shown below.
7. Insert the tubing into the center port of the Multigas assembly; use the nut to secure the
tubing to the assembly.
8. Insert the other end of the tubing into the gas inlet on the rear panel of the analyzer; use the
self-contained swage nut to secure the tubing to the inlet.
Ferrules
Swage nut
Ferrules
Swage nut
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...