AccuPyc II 1340
Installing the Analyzer
June 2013
2-11
Calibrating the Pycnometer
Although calibrated before shipment, the pycnometer should be recalibrated to your laboratory
atmospheric and environmental conditions. Calibration consists of a two-part procedure:
• Specifying calibration parameters
• Performing the calibration
Specifying Calibration Parameters
Calibration parameters are specified using the Analysis Parameters prompts.
Refer to the following sections if you wish to familiarize yourself with the keypad and display
before proceeding:
•
•
1.
Press
ALT + 2
to access the
Set up type
prompt.
Analysis Parameters
should display as the Set-up type; if not, press
CHOICE
until it
displays.
2.
Press
ENTER
to accept
Analysis Parameters
and advance to the next prompt; accept the
default value of
10
for the number of purges.
3.
Press
ENTER
; accept the default of
19.500 psig
for the fill pressure.
Refer to
if you need additional information on responses to the
prompts in this procedure.
Unit[n]
SN1234 10 cm3
Set-up type?
Analysis Parameters
Unit[n]
SN1234 10 cm3
Analysis Parameters
Number of purges:
10
Unit[n]
SN1234 10 cm3
Analysis Parameters
Purge fill pressure:
19.500 psig
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...