AccuPyc II 1340
Appendix J
June 2013
J-5
4. Place the cup into the insert, then tilt the tool to one side to remove it from the sample cup.
5. If using a fritted filter cap, place it onto the top of the insert.
6. Replace the sample chamber cap.
After the operation is complete, perform the following steps to remove the sample cup and
insert:
1. Remove the sample chamber cap and place on a clean work surface with the greased side
facing upward.
2. Remove the fritted filter cap (if used).
3. Insert the handling tool into the sample cup so that it is gripped by the O-ring.
4. Lift the sample cup out of the insert.
5. Use your fingers to pull the insert from the sample chamber.
6. Replace the sample chamber cap.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...