Maintenance
AccuPyc II 1340
5-4
Aug 2013
Maintenance
Calibrating the Pycnometer
You should check or calibrate the pycnometer every time you restart it. For a quick check, run
an analysis with an empty cup to see how close the average volume is to 0. It should be
±0.05% of full-scale. If it is not within ± 0.05% of full-scale, calibrate according to the
instructions provided in
pycnometer, cell volume and expansion volume in the set-up parameters are updated
automatically.
Greasing the Chamber Cap O-Ring
The cell chamber cap contains an O-ring that requires routine maintenance because it is so
often exposed to atmospheric conditions. The chamber cap O-ring should be greased at the
beginning of each period of use. Wear rubber gloves to prevent contaminating the grease with
oil from your fingertips.
1.
Turn the chamber cap counterclockwise and lift it from the chamber.
2.
Place the chamber cap on a clean surface, with the O-ring side exposed.
3.
Place a small amount of Dow Corning High vacuum grease (or equivalent) on your index
finger.
4.
Run your index finger around the O-ring groove.
5.
Replace the chamber cap.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...