System Commands
AccuPyc II 1340
3-12
Aug 2013
pycnometer is shipped with default values; however, you may wish to modify them to meet your labo-
ratory requirements.
Press
ENTER
to accept
Analysis Parameters
and display the
next prompt.
Enter the number of purges to be performed and press
ENTER
.
Range:
0 to 999
Purging cleans the sample cell and expansion chambers
before an analysis begins. The greater the number of purges,
the cleaner the sample will be when analyzed.
For
calibration
,
10
is sufficient.
Enter the fill pressure and press
ENTER
. For most
applications, the default of 19.500 psig is adequate.
Range:
0 to 19.850 psig
0 to 136.86 kPag
Typically, the greater the fill pressure, the easier it is to
measure the volume precisely. However, a lower pressure
may be desirable for some samples.
Enter the number of cycles to be performed and press
ENTER
. A cycle is a series of functions which equal to a
single volume measurement.
Range:
1 to 999
For
calibration
,
10
is sufficient.
Enter the fill pressure and press
ENTER
. For most
applications, the default of 19.500 psig is adequate.
Range:
0 to 19.850 psig
0 to 136.86 kPag
Unit[n]
SN1234 10 cm3
Set-up Type?
Analysis Parameters
Unit[n]
SN1234 10 cm3
Analysis Parameters
Number of Purges:
10
Unit[n]
SN1234 10 cm3
Analysis Parameters
Purge Fill Pressure:
19.500 psig
Unit[n]
SN1234 10 cm3
Analysis Parameters
Number of cycles:
10
Unit[n]
SN1234 10 cm3
Analysis Parameters
Cycle Fill Pressure:
19.500 psig
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...