AccuPyc II 1340
Troubleshooting
Aug 2013
5-3
Unit will not equilibrate,
or results are not
reproducible.
Sample outgassing.
Prior to analysis, remove moisture and
contaminants from sample. (Refer to
Preparing and Loading a Sample
.)
Pycnometer leaks.
Check cap O-ring for defects. Regrease
or replace the O-ring, if necessary.
(Refer to
.)
While in Manual Mode, with the cap
installed, open all valves. With helium
flowing through the pycnometer, actuate
the expansion valve repeatedly for
approximately 1 minute. This will
remove lint, which could cause a leak,
from the valve seat. Repeat the process
for the vent valve with the inlet and
expansion valves open.
Dust filter on 2000 cm
3
unit is clogged or
leaking to atmosphere.
Clean the dust filter. Refer to
the Dust Filter (2000-cm3 units only)
.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...