AccuPyc II 1340
System Commands
Aug 2013
3-11
•
Multiple choice
prompts contain a fixed set of responses and are followed by a
question mark (?). To select a multiple-choice response, press
CHOICE
until the desired
response is displayed, then press
ENTER
.
At any time while entering information you may:
• Press
SAVE
to save the information you entered and return to the
Reload
prompt.
• Press
Alt + CLEAR
to discard the information you entered and return to the
Reload
prompt.
Set Up
Press
Alt
+
2
to access the set-up function.
At the completion of each set-up, you are returned to the
Set-up type?
prompt. Choose one of the fol-
lowing options:
• Press
SAVE
to save the information you entered and return to the
Reload
prompt.
• Choose another set-up type; do not press
SAVE
unless you are through entering parameters
and wish to return to the
Reload
prompt.
• Press
Alt
+
CLEAR
to discard the information you entered and return to the
Reload
prompt.
Analysis (and Calibration) Parameters
This option enables you to specify analysis and calibration parameters. Although the
Set-up Type
is
shown as Analysis Parameters, these same prompts are used to specify Calibration Parameters. The
Set Up
enables you to choose or enter parameters to be used
for analysis, calibration, reporting, and data transmission.
Choices:
Analysis Parameters, Report Options,
Calibration Data, Communications, Unit Types
Press
CHOICE
to display the desired set-up type, then press
ENTER
.
The parameters you enter or choose for each set-up type are
stored and used to control the pycnometer until new ones are
specified. Each
Set Up
command is explained in subsequent
sections.
Unit[n]
SN1234 10 cm3
Set-up type?
Analysis Parameters
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...