Viewing and Printing Data Results
AccuPyc II 1340
4-6
June 2013
Viewing and Printing Data Results
After analysis or calibration, the display returns to the
Reload
prompt.
A report is generated automatically to the destination you select at the
Set Up > Report
Options
as follows:
•
Printer
: report is automatically sent to the printer
•
Transmission line
: report is automatically sent through the serial line to an attached
device for this purpose
•
: report is e-mailed to the address you specified (when connected to a network)
An asterisk at the
Reload
prompt indicates that error messages have been generated during
operation. (Refer to
for additional information on viewing
and deleting error messages.) If you wish to view report data by points, use the
Review
func-
tion explained on page
View Data
Reports are generated after analysis and calibration, and remain available for viewing or print-
ing until another automatic operation (other than zero) is performed. When you perform an
automatic operation, data from the previous operation are no longer available for viewing on
the display.
A display report is always generated regardless of the specified destination and contains data
as follows:
•
Analysis
: the average density (or volume) and the deviation from the mean.
•
Calibration
: average cell (or expansion volume) and the deviation from the mean
Print Data
To retain a record of analysis or calibration data, press
Alt + 6
to generate a report to an
attached printer. Printed reports contain data as follows:
•
Analysis
: volume, density, and deviation for each cycle, as well as an average of all
cycles. The report also contains the date and time the analysis was started and
completed and the temperature of the cell chamber.
•
Calibration
: cell and expansion volumes and deviations for each cycle, as well as an
average of all cycles. The report also contains the date and time the calibration started
and completed.
Unit[n]
SN1234
10 cm3
X - X - X
Reload *
P = (current pressure)
An asterisk (*) indicates
error message(s) exist
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...