AccuPyc II 1340
Appendix A
Oct 2013
A-11
USR_ERR:
No data to review
USR_ERR: Number of
cycles must be = 5
USR_ERR: Out of range
USR_ERR: Pressure overrange
Valve failure
[ENTER] to retry.
Cause:
You tried to review data for an automatic operation when there were no
data to review.
Action:
Abandon request.
Cause:
You tried to enable run precision without increasing the number of runs
to at least five.
Action:
Increase the number of runs to five or abandon request.
Cause:
You tried to enter a value that is out of the valid range.
Action:
Enter a value in the specified range (refer to Chapter 4).
Cause:
A pressure overrange occurred but was left uncorrected.
Action:
Return to manual mode and vent the system.
Cause:
One (or more) of the system valves failed to operate when the
instrument was initialized.
Action:
Press
ENTER
to retry. If still unsucessful, contact your service
representative.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...