Equipment Description
Aug 2013
1-5
designated purge fill pressure. The chambers are then vented to atmosphere, resulting in elimination of
water vapor or other contaminants. A
cycle
is used for collecting the precise, accurate data used in
report calculations.
Calibration
is used to determine the size of the cell and expansion chambers within the instrument.
After calibration, the cell and expansion chamber volumes are stored automatically.
The
zero
function is used to calculate a new zero pressure offset. It is measured when the system is at
atmospheric pressure and thermal equilibrium and is subtracted from all subsequent pressure readings
in order to obtain a true gauge pressure.
Configurations
To ensure best fit with your sample, the AccuPyc is available in multiple configurations:
• 1-cm
3
sample chamber
• 10-cm
3
sample chamber
• 100-cm
3
sample chamber
• 350-cm
3
sample chamber
• 2000-cm
3
sample chamber
Best fit means your sample nearly fills the sample chamber and, therefore, optimizes the precision of
your results.
Temperature Control
A temperature
-
controlled unit to which an external bath is connected is available in the following
configurations:
• 10-cm
3
sample chamber
• 100-cm
3
sample chamber
The temperature-controlled unit is specifically designed for temperature-sensitive materials. This unit
is ideally suited for laboratories in which ambient temperature varies during normal work hours, or in
which subambient cooling or above-ambient heating is required.
All sample chamber volumes are nominal unless otherwise specified.
Содержание AccuPyc 2 1340
Страница 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Страница 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Страница 4: ......
Страница 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Страница 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Страница 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Страница 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Страница 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Страница 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Страница 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Страница 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Страница 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Страница 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Страница 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Страница 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...
Страница 158: ...Index AccuPyc II 1340 Index 6 June 2013...