GETTING STARTED
Logicube OmniClone® Xi User’s Manual
10
Touch Screen
The OmniClone Xi features an LCD Touch Screen
that allows the user to quickly input commands. This
screen replaces many of the buttons that were
present on older Logicube duplication products. The
screen is bright and easy to read. It also emits an
audible beep every time the touch screen is tapped.
This lets the user know that the touch screen is active.
Calibrating the Touch Screen
There may be times when the user wants to
recalibrate the Touch Screen. The procedure for this
is very simple as outlined in the steps below:
1. Press the Reset button on the back of the unit and
hold it in.
2. Press and hold the Set button, then release the
Reset button.
3. Hold the Set button until the unit boots to a screen
that reads “Touchpad Calibration. Please touch
square with fi
nger (1/5)”.
4. Look for a square at the top of the screen. Touch
or tap the square when it is located.
5. Repeat the previous step four more times. The
unit will count each time the square is tapped
correctly. It will count (1/5), (2/5), etc.
6. Once the screen has been calibrated, the contrast
adjustment screen will appear. Adjust the contrast
as needed, then tap the
Back
icon. The
OmniClone Xi will return to the Main Menu
Screen.
Buttons
The OmniClone Xi
features two “Shortcut” buttons
that are located to the right of the touch Screen.
These buttons are available at all times.
START/STOP
Button
– Press it twice to begin a
cloning operation using the current settings; press
the START/STOP button in mid process to abort it.
A single key stroke presents a preview screen
where you can see the current setting, and decide
whether to press it again to begin the capture, or
back out to reconfigure.
SET
Button
– Press this button once to bring
you to the settings screen where you can change
capture modes and other settings of the unit.