Troubleshooting
4
Infiniium V-Series Oscilloscopes Service Guide
61
The first set of LEDs is located in the upper right corner of the backplane assembly
(see the following two figures). Six of the LEDs indicate an under-voltage fault,
and four indicate an over-temperature fault. Another LED in this group is BULK
SUPPLY FAIL, located below NEG SUPPLY FAIL. It lights up if the bulk supply needs
replacing.
Under-voltage, over-temperature, and bulk supply indicators on backplane assembly