Troubleshooting
4
Infiniium V-Series Oscilloscopes Service Guide
81
MSO Assembly Troubleshooting
This section describes problems that may occur when an MSO assembly is
installed, and how to troubleshoot them. The MSO printed circuit assembly is
located below the lower acquisition board, with the components facing downward.
LEDs on MSO assemblies
The MSO printed circuit board has two sets of LEDs. One set is visible from the
right side of the oscilloscope, and the other set is visible from the left side.
MSO LEDs on the right side of the oscilloscope
To see the LEDs on the right
side, look straight in through the vent holes and see the lights pointed toward the
bottom of the oscilloscope. These two red LEDs turn on when the oscilloscope is
turned on, indicating that the comparators are off. The LEDs stay on until the
FPGA is programmed by the scope application software; then the software turns
the LEDs off when it turns the comparators on.
If the LEDs turn on after the application is running, a thermal problem likely was
detected. A warning appears. To resolve the problem, shut down the scope and
allow it to cool. Make sure the fan holes are not blocked, and allow at least 4
inches of space next to the oscilloscope. Restart the scope and verify that all fans
are working.
MSO LEDs on the left side of the oscilloscope
The LEDs on the left side can
be seen only as a reflection in the lower sheet metal as you look through the fans.
These LEDs are off until the FPGA is programmed successfully, which occurs
shortly after the scope application starts.
One of the LEDs turns on and remains on after the FPGA is programmed. The other
LED will start blinking steadily at about 1.6 Hz if the 250 MHz clock coming from
the backplane board is working properly. If the LED is not blinking steadily, verify
that the backplane board was modified correctly to be used with the MSO board.
That is, in the clock area on the top side of the backplane board, R352 and R353
should be loaded with 0 ohm resistors. The same area on the bottom side of the
board should not have resistors on R330 and R332. If this clock rework has not
been done, you will see an error message indicating an option card DAC failure
when the application starts up.
Содержание Infiniium MSOV 084A
Страница 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Страница 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Страница 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Страница 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Страница 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Страница 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Страница 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Страница 163: ......