18
Infiniium V-Series Oscilloscopes Service Guide
2
Calibration
4
Click
Start Quick Env Cal
, then follow the instructions on the screen. The routine
will prompt you to follow these steps:
a
Disconnect everything from all inputs and Cal Out.
b
Connect the calibration cable from Cal Out to channel 1.You must use the
54916-61626 cable assembly with two connector saver adapters. Failure to
use the appropriate calibration cable will result in an inaccurate calibration.
c
Connect the calibration cable from Cal Out to each channel input as
requested.
d
A Passed/Failed indication appears for each calibration section. If any
section fails, click the Enable Details box for information on the failures. Also
check the calibration cables.
Clear this check box
before beginning
calibration
Click here to run a
quick environmental
calibration