Troubleshooting
4
Infiniium V-Series Oscilloscopes Service Guide
63
Locate the other two groups of over-voltage fault LEDs along the edges of the two
acquisition assemblies. They are visible through the bottom fan opening located
on the left side of the oscilloscope, nearest the rear panel (see the following
figures).
Over-voltage indicator LEDs on acquisition assembly behind fan
Over-voltage indicator LEDs with fan removed
The acquisition assembly nearest the bottom of the oscilloscope is the channels 1
and 2 acquisition assembly and the one above it is the channels 3 and 4
acquisition assembly. If you look through the fan and see any of the over-voltage
LEDs lit, that acquisition assembly needs to be replaced.