44
Infiniium V-Series Oscilloscopes Service Guide
3
Testing Performance
Time Scale Accuracy (TSA) Test
This procedure verifies the maximum TSA specification for the oscilloscope.
Description
TSA refers to the absolute accuracy of an oscilloscope’s time scale. Because TSA
depends directly on frequency of a crystal oscillator, it is comprised of two
components: an initial accuracy component and an aging component.
The initial accuracy component applies to the oscilloscope’s accuracy immediately
after a time base calibration, whether performed at the factory, by a customer, or
by a Keysight service center. The aging component scales linearly from the time
since the last time base calibration, and adds to the initial accuracy component.
Equipment Required
Description
Critical specifications
Recommended model/part #
Synthesized sine wave
source
Output Frequency:
≥
10 MHz
Output Amplitude: 0 dBm
Frequency Resolution: 0.1 Hz
E8257D PSG
10 MHz frequency
reference
Output Frequency: 10 MHz
Absolute Freq. Error: <
±
0.0275 ppm
53132A opt. 012 frequency
counter
RF cable
50
Ω
characteristic impedance (no
substitute)
BNC (m) connectors
Max Frequency:
≥
50 MHz
8120-1840
Adapters, assorted
3.5 mm (f) to Precision BNC (m)
3.5 mm (f) to 3.5 mm (f)
54855-67604
83059B
Содержание Infiniium MSOV 084A
Страница 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Страница 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Страница 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Страница 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Страница 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Страница 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Страница 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Страница 163: ......