Replacing Assemblies
5
Infiniium V-Series Oscilloscopes Service Guide
131
Removing and replacing the optional GPIB port
1
Disconnect the power cable and remove the top panel.
2
Remove the screw from the top edge of the chassis.
3
Pull the port out of the slot.
)
To replace the standard profile bracket with a V-Series Profile Bracket:
1
Remove the hex screws that hold the connector bracket and the standard
profile bracket by using a hex 9/32”, tightening torque 7.0 in/lbs screwdriver.
2
Replace the standard profile bracket with the V-Series Profile Bracket and
replace the screws.
Figure 78
NOTE
If you are replacing the GPIB port with a new one (82351B or newer), you will need to use a
V-Series Profile Bracket (54919-00105).