124
Infiniium V-Series Oscilloscopes Service Guide
5
Replacing Assemblies
5
Slide the heat shrink with the cable part number toward the SMPM connector
as far as it will go for cable input.
)
6
Carefully remove the input cables from the old acquisition board.
7
Insert the right-angle SMPM end of the first input cable through the bottom of
the acquisition tray assembly. The cables route through the plastic grommet
and through the oval hole in the acquisition board as shown.
)
Figure 67
CAUTION
For the following steps, be careful not to crimp or pinch the input cables. Also, to
ensure the weight of the assembly does not rest on the cables, support the tray
from underneath.
Figure 68