Replacing Assemblies
5
Infiniium V-Series Oscilloscopes Service Guide
111
11
Remove the two M4 screws from the top of the backplane board and two M4
screws from the rear of the oscilloscope.
)
12
Ensure all cables are disconnected from the backplane board. Look from both
the top and bottom of the oscilloscope to see and reach them all.
13
Lift the entire acquisition/backplane assembly out of the chassis.
))
Figure 44
Figure 45