92
Infiniium V-Series Oscilloscopes Service Guide
5
Replacing Assemblies
6
Remove the two screws from the front panel deck.
7
Turn the oscilloscope upside down and remove the three M3 screws from the
rear of the bottom panel. Remove the remaining 13 screws from the bottom
panel and lift the bottom panel off the chassis.
8
Disconnect the two Cal Out/Aux Out cables from the backplane board. Pull the
cables straight back to remove them from the backplane board.
Figure 8
CAUTION
Be sure to pull the cables straight back to avoid bending or breaking the connection to
the backplane board.