Testing Performance
3
Infiniium V-Series Oscilloscopes Service Guide
27
Zero Error Test Procedure
1
Disconnect all cables from the oscilloscope channel inputs.
2
Click
Control > Factory Default
.
3
Click
Setup > Acquisition...
. In the Acquisition dialog box, enable averaging and
set # of Averages to 256.
4
Configure the oscilloscope to measure the average voltage (V avg) on channel
1 as follows:
a
Change the vertical scale of channel 1 to 10 mV/div.
b
Click the Vertical Meas tab on the left side of the waveform window, then
drag and drop the Average measurement icon onto the channel 1 waveform.