46
Infiniium V-Series Oscilloscopes Service Guide
3
Testing Performance
13
Record the time since calibration (in years) in the table. The calibration date
can be found in the Calibration window (
Utilities > Calibration...
).
14
Calculate the test limits using the following formula and record them in the
table.
Test Limits = ±(0.100 + 0.100 x Years Since Calibration)
15
Record the results in the Performance Test Record.