Troubleshooting
4
Infiniium V-Series Oscilloscopes Service Guide
73
Motherboard Verification
If you have been through the Power Supply Troubleshooting section of this chapter
and the oscilloscope still does not stay powered up, the problem may be with the
motherboard. To diagnose this problem, follow these steps.
1
Remove the top panel.
2
Turn the oscilloscope on.
If the unit beeps once after about 5 to 10 seconds but fails to boot up, the
motherboard is defective and needs to be replaced.
If the unit fails to beep after it is powered on, then measure the potential
relative to the chassis at pin 9 of J1201 on the backplane board (see figure
below). Pin 9 is the reset line for the motherboard and is controlled by the
backplane. You should measure about 3.3 V. If the potential you measure is
correct, but the unit still has not beeped and started up, the motherboard is
defective and needs to be replaced.
3
If the potential you measure is not correct, shut the oscilloscope down and
unplug the AC cord from the rear of the instrument.
4
Unplug the front panel power cable, the two large power cables that go to the
motherboard (J1201 and J1202), and the red and blue SATA cables (J900 and
J901) from the PCIe-to-SATA adapter board on the backplane assembly. Do not
allow the two SATA cables to touch the backplane assembly and cause a short
circuit.
5
Plug the AC power back into the oscilloscope and turn the unit on if it does not
come on automatically.
6
Measure the potential relative to the chassis at pin 9 of J1201 again.
If you measure the required 3.3 V, the motherboard is defective and needs to be
replaced.
If you did not measure the correct voltage, the backplane assembly is defective
and must be replaced.
Measure potential relative to chassis at pin 9 of J1201
Содержание Infiniium MSOV 084A
Страница 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Страница 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Страница 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Страница 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Страница 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Страница 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Страница 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Страница 163: ......