30
Infiniium V-Series Oscilloscopes Service Guide
3
Testing Performance
Offset Gain Test Procedure
1
Make the connections to oscilloscope channel 1 as shown below.
2
Set up the DMM to perform DC voltage measurements.
3
Click
Control > Factory Default
.
Oscilloscope channel input
Cal Out on oscilloscope front panel
BNC (f) to SMA (m) adapter
BNC (f) to SMA (m) adapter
Connector saver
Connector saver
NOTE
•
Where the BNC Tee adapter is used, it is important to connect it directly to the oscilloscope
channel input using the BNC (f) to SMA (m) adapter and the connector savers to minimize
ground potential differences and to ensure that the DMM measures the input voltage to the
oscilloscope channel as accurately as possible. Differences in ground potential can be a
significant source of measurement error, particularly at high scope sensitivities.
•
It also helps to reduce ground potential differences if the oscilloscope and DMM are
connected to the same AC supply circuit.
•
256 averages are used in the oscilloscope measurements of this section to reduce
measurement noise and to reduce the measurement error due to resolution.
Содержание Infiniium MSOV 084A
Страница 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Страница 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Страница 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Страница 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Страница 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Страница 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Страница 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Страница 163: ......