Testing Performance
3
Infiniium V-Series Oscilloscopes Service Guide
47
Performance Test Record
Keysight Technologies
Keysight Infiniium V-Series Oscilloscopes
Model Number _____________________
Tested by________________________
Serial Number ___________________________
Work Order No.___________________
Recommended Test Interval—1 Year/2000 hours
Date_____________________________
Recommended next test date ___________________
Ambient temperature ______________
Offset Accuracy—Zero Error Test
Vertical Scale
Test Limits
Channel 1
Channel 2
Channel 3
Channel 4
10 mV/div
-1.8 mV to +1.8 mV
20 mV/div
-2.6 mV to +2.6 mV
50 mV/div
-5.0 mV to +5.0 mV
100 mV/div
-9.0 mV to +9.0 mV
200 mV/div
-17.0 mV to +17.0 mV
500 mV/div
-41.0 mV to +41.0 mV
1 V/div
-81.0 mV to +81.0 mV