Replacing Assemblies
5
Infiniium V-Series Oscilloscopes Service Guide
95
13
Pull up and tilt the front panel to see the cables connected to it. Route the
display cable and on/off cable through the grommet on the chassis after
disconnecting them from the other side of the connections.
14
Remove the front panel.
To replace the front panel assembly, reverse this procedure.
Figure 14
CAUTION
When setting the front panel on the chassis to install the front panel assembly onto
the oscilloscope, make sure not to damage the cables.