Testing Performance
3
Infiniium V-Series Oscilloscopes Service Guide
45
Connections
Connect the equipment as shown here.
Procedure
1
Configure the sine wave source to output a 0 dBm (600 mVpp) sine wave into
50 ohms with a frequency of 10.00002000 MHz.
2
Adjust the source amplitude such that the displayed sine wave is 600 mVpp.
3
Click
Control > Factory Default
.
4
Set the vertical scale of channel 1 to 100 mV/div.
5
Set the oscilloscope’s sampling rate to 100 kSa/s.
6
Set the oscilloscope's horizontal scale to 20 ms/div.
7
Set the measurement thresholds for all waveforms to a fixed voltage level of 0 V
and ±20 mV hysteresis:
a
Click
Measure > Thresholds...
.
b
Select
Custom: level +/- hysteresis
from the Thresholds drop-down list box.
c
Enter 20 mV in the
Hysteresis
field and 0 V in the
Threshold Level
field.
8
Enable a frequency measurement on channel 1.
9
On the oscilloscope, press
[Stop]
.
10
Press
[Clear Display]
.
11
Press
[Run]
, wait until 10 acquisitions have accumulated, and then press
[Stop]
.
12
Convert the average frequency value to time scale error by subtracting 20 Hz
and dividing by 10 Hz/ppm. Record the result in the Measured Time Scale Error
(ppm) column of the Time Scale Accuracy table.
Содержание Infiniium MSOV 084A
Страница 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Страница 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Страница 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Страница 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Страница 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Страница 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Страница 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Страница 163: ......