How Supplied
The EndoVac
™
Complete Installation Kit (973-3003) includes all major components*:
A.
Handpiece (2)
H. 3 cc Syringes for EDTA (6)
B.
Fingerpiece (2)
I.
MDT Evacuation Tubing
(Blue Connectors) (2)
C.
Multi-Port Adapter (1)
D.
MacroCannulas 25mm (20)
J.
Handpiece/Fingerpiece Evacuation Tubing
(White Connectors) (2)
E.
MicroCannulas** 25mm (20)
F.
20 cc Syringes for NaOCl (6)
K.
Instructions for Use (not shown)
G.
Master Delivery Tips MDT(20)
L.
Quick Setup Guide (not shown)
C
A
D
E
B
F
H
I
J
G
*Most components above can also be re-ordered separately.
**MicroCannulas can be reordered in 21, 25, 31 mm and 50 mm lengths.
Indications for Use
The EndoVac
™
system is intended for the delivery and evacuation of endodontic irrigation solutions during
root canal procedures.
Contraindication
None Known
Description of Device
The EndoVac
™
System is intended for the delivery and evacuation of endodontic irrigation solutions during
root canal procedures. EndoVac
™
Apical Negative Pressure Irrigation System is used in root canal therapy and
is comprised of four main components. The Multi-Port Adapter is used to connect the Master Delivery Tip to
the dental office Hi-Vac system. The Master Delivery Tip provides a constant flow of irrigant that lowers the
risk of overflow due to continuous suction. Additional tips include the MacroCannula which is used to remove
coarse debris after instrumentation, and the MicroCannula which features microscopic evacuation holes
allowing the irrigant to flow to the apical termination.
Warnings & Precautions
•
Prior to EndoVac
™
irrigation, the clinician must take every precaution necessary to ensure the integrity
of the rubber dam seal and any potential communication from the pulp chamber to the oral cavity are
absolutely secure.
•
Always protect the patient’s eyes with safety glasses and clothing from sodium hypochlorite splatter or
spill.
•
Never place the MDT’s delivery tip closer than five (5) millimeters from the coronal opening of any pulp
canal.
•
Never aim the MDT directly into the orifice of the canal as this may create a positive force of irrigant into
the canal.
•
In order for the MicroCannula to be placed to full working length (WL), it requires a minimum canal
preparation size of at least a size #35 /.04 taper, to full Working Length (WL) or at least a size #45 when
a non-tapered instrument, such LightSpeed
®
LSX
™
, is used.
•
Always use the EndoVac
™
system in the sequence described herein. Skipping or deviating from the
following steps can cause the EndoVac
™
cannulae to clog.
•
Pre-fill various syringes: 20 cc syringes with sodium hypochlorite, 3 cc syringes with EDTA, and optional
syringes with irrigants of the clinician’s choice. Make sure no air bubbles are trapped in the pre-filled
syringes, as this will cause uncontrolled irrigant extrusion after releasing the plunger pressure.
Adverse Reactions
None Known
Содержание EndoVac
Страница 1: ...EndoVac Apical Negative Pressure Irrigation System Instructions for Use...
Страница 3: ...3 1 Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7...
Страница 4: ...4 2 3 Figure 8 Figure 9 Figure 10 Figure 11 Figure 12 Figure 13...
Страница 5: ...5 4 5 Figure 14 Figure 15 Figure 16 Figure 17 Figure 18 Figure 19...
Страница 6: ...6 6 7 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 Figure 25...
Страница 67: ...67 MDT 10 MDT Micro MacroCannnula 15 3 FDA 132 C 4 20 134 C 3 10 10 SybronEndo EndoVac 12 SybronEndo...