4-14
Return to
2600AS-901-01 Rev. B / September 2008
Section 4: Source-Measure Concepts
Series 2600A System SourceMeter® Instruments Reference Manual
Figure 4-8
V-Source operating examples
Current Limit
Load Line
Voltage Source
Load Line
V-Source (V
S
)
I-Meter
(I
M
)
5mA
10mA
10V
2k
W
DUT Load Line (R)
I
M
= V
S
/ R
V
S
= I
M
R
= 10V/2k
W
= 5mA
A) Normal V-source operation
Current
Limit
Load Line
Operating
Point
Voltage
Source
I-Meter
(I
M
)
10mA
10V
= (10mA) (800
W
)
= 8V
B) V-Source in compliance
8V
800
W
DUT Load Line (R)
V-Source (V
S
)
Operating
Point