19-256
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
Parameters
smu,bias,start,stop,limit,ton,toff,points,buffer,tag
[,sync_in][,sync_out])
smu
:
SourceMeter instrument channel (for example,
smua
).
bias:
Pulse bias level in amps.
start:
Pulse sweep start level in amps.
stop:
Pulse sweep stop level in amps.
limit:
Voltage limit (for example, compliance) in volts.
ton:
Pulse width (for example, on-time) in seconds.
toff:
Pulse off time in seconds.
points:
Number of pulse measure cycles.
buffer:
Reading buffer where pulsed measurements will be stored.
If this is
nil
when the function is called, no measurements
will be made when the pulse train is initiated.
tag:
Numeric identifier to be assigned to the defined pulse train.
sync_in:
Defines a digital I/O trigger input line. If programmed, the
pulse train will wait for a trigger input before executing each
pulse. This parameter is optional.
sync_out:
Defines a digital I/O trigger output line. If programmed, the
pulse train will generate a trigger output immediately prior
to the start of
ton
. This parameter is optional.
sync_in_timeout:
Specifies the length of time (in seconds) to wait for input
trigger (defaults to 10s).
New to Firmware 1.3.0
sync_in_abort:
(true or false) Whether to abort pulse if input trigger is not
received before
timeout
expires (defaults to
true
).
New
to Firmware 1.3.0
Waveform
Table 19-14 (cont.)
KIPulse TSP test script: ConfigPulseIMeasureVSweepLin
TSP project name:
KIPulse
TSP test script name:
KIPulse
Function:
ConfigPulseIMeasureVSweepLin
Usage:
f, msg = ConfigPulseIMeasureVSweepLin
(smu,bias,start,stop,limit,ton,toff,points,buffer,tag[,sync_in]
[,sync_out][,sync_in_timeout][,sync_in_abort])
Pul
s
e
initi
a
ted
h
ere
bias
le
v
el
toff
start
Me
as
ure
m
ent
(
1/NPLC
s
econd
s)
step
stop
ton
step = (stop - start)/(points -1)
L
as
t
s
ource
va
lue
b
e
f
ore
p
ul
s
e
tr
a
in
was
initi
a
ted
step