2600AS-901-01 Rev. B / September 2008
Return to
16-9
Series 2600A System SourceMeter® Instruments Reference Manual
Section 16: LAN Concepts and Settings
Assigning the Method
Complete the following steps to select a method:
1.
From the front panel, press
MENU
>
LAN
>
CONFIG
>
METHOD
.
2.
Select one of the following methods and then push the
navigation wheel
to enter the
desired method:
•
Press
AUTO.
•
Press
MANUAL.
The LAN CONFIG menu is shown.
3.
(Optional) Press
Exit
to return to the LAN CONFIG menu. CONFIG blinks on the front
panel.
4.
Turn the navigation wheel one click to the left and then push-in to select
APPLY_SETTINGS
.
5.
Select
YES
.
Setting the IP address
Note the following:
•
Contact your IT department to secure a valid IP address for the instrument when placing
the instrument on a corporate network.
•
The IP address does not need to be manually set if the method is set to AUTO.
A direct connection to the PC can also be configured. See
Establishing a point-to-point connection
for more information.
Setting the subnet mask
Note the following:
•
Contact your IT department to secure a valid subnet mask for the instrument when
placing the instrument on a corporate network.
•
The subnet mask does not need to be manually set if the method is set to AUTO.
A direct connection to the PC can also be configured.
Establishing a point-to-point connection
for more information.
Understanding the domain name system
The Domain Name System (DNS) lets you type a domain name in the address bar to connect to
the instrument. The DNS removes the requirement to memorize the IP address, instead you only
need to know the domain name.
Example
:
Series2600A.yourcompany.com
Contact your IT department to learn more about DNS.
NOTE If a DNS server is not part of the LAN infrastructure, then this setting
is not used.