Series 2600A System SourceMeter® Instruments Reference Manual
tspnet.tsp.abortonconnect
Attribute
Abort on connect.
Default
1
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
tspnet.tsp.abortonconnect = value
value = tspnet.tsp.abortonconnect
value
1
or
0
Remarks
• This setting determines if the Series 2600A sends an abort message when it attempts to connect
using
TM
-enabled instrument.
• Sending the
abort
command on a connection causes any other active interfaces being used on
that instrument to close to ensure you have obtained access to the remote instrument.
• Connecting to a TSP instrument without issuing an
abort
command, or when
tspnet.tsp.abortonconnect
is set to 0, can result in the Series 2600A suspending
operation until it receives a response back from the instrument or until a timeout error is
generated.
Example
tspnet.tsp.abortonconnect = 0