2600AS-901-01 Rev. B / September 2008
Return to
22-3
Series 2600A System SourceMeter® Instruments Reference Manual
Section 22: Performance Verification
Line power
The Series 2600A requires a line voltage of 100V to 240V and a line frequency of 50Hz or 60Hz.
Verification tests should be performed within this range.
Recommended test equipment
summarizes recommended verification equipment. You can use alternate equipment as
long as that equipment has specifications equal to or greater than those listed in
. Keep
in mind, however, that test equipment uncertainty will add to the uncertainty of each measurement.
Generally, test equipment uncertainty should be at least four times better than corresponding
Series 2600A specifications.
lists the uncertainties of the recommended test
equipment.
Verification limits
The verification limits stated in this section have been calculated using only the Series 2600A
one-year accuracy specifications, and they do not include test equipment uncertainty. If a
Table 22-1
Recommended verification equipment
Description
Manufacturer/Model
Accuracy
Digital Multimeter
Keithley Instruments
Model 2002
or
Agilent 3458A
DC Voltage
1
(2601A/2602A)
DC Voltage
2
(2611A/2612A/
2635A/2636A)
DC current
3
90mV:
0.9V:
5.4V:
36V:
190mV:
1.8V:
18V:
180V:
90nA:
0.9µA:
9µA:
90µA:
0.9mA:
9mA:
90mA:
0.9A:
±8ppm
±5ppm
±4ppm
±6ppm
±5ppm
±4ppm
±6ppm
±6ppm
±430ppm
±45ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
0.5
Ω
, 250W, 0.1%
Precision Resistor
Isotek
RUG-Z-R500-0.1-TK3
Resistance
4
0.5
Ω
:
±125ppm
1G
Ω
, 200V, 1%
standard
Keithley Instruments
Model 2600-STD-RES
Resistance
5
1G
Ω
:
250ppm
1. 90-day specifications show full-range accuracy of recommended model used for specified
measurement point.
2. Id.
3. Id.
4. Resistor used to test Model 2601A/2602A 3A range and Model 2611A/2612A/2635A/2636A
1.5A range only should be characterized to uncertainty shown using resistance function of
digital multimeter before use.
5. Standard is a guarded and characterized 1 G
Ω
resistor used to test Model 2635A/2636A
100pA to 100nA current ranges.