Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
Factory script information
KISweep
The KISweep factory script provides simple sweep test programming and can be used as an
example of how to use the sweeping function.
The KISweep factory script is made up of the following functions. All of these functions can be
accessed from both the front panel and the remote interfaces. These functions provide simple
sweeping operations and can be used as examples of how to use the sweeping functions. Details
on these functions are provided in the tables that follow.
SweepILinMeasureV(smu, starti, stopi, stime, points)
SweepVLinMeasureI(smu, startv, stopv, stime, points)
SweepILogMeasureV(smu, starti, stopi, stime, points)
SweepVLogMeasureI(smu, startv, stopv, stime, points)
SweepIListMeasureV(smu, ilist, stime, points)
SweepVListMeasureI(smu, vlist, stime, points)
Table 19-3
KISweep TSP test script: SweepILinMeasureV
TSP project name:
KISweep
TSP test script name:
KISweep
Function:
SweepILinMeasureV(smu, starti, stopi, stime, points)
Description
Performs a linear current sweep with voltage measured at every step (point):
• Sets the
smu
to output
starti
amps, allows the source to settle for
stime
seconds, and then performs a voltage measurement.
• Sets the
smu
to output the next amps step, allows the source to settle for
stime
seconds, and then performs a voltage measurement.
• Repeats the above sequence until the voltage is measured on the
stopi
amps step.
The linear step size is automatically calculated as follows:
step = (stopi – starti) / (points – 1)
starti
=
Volt
ag
e
m
e
as
ure
m
ent
stopi
stime
stime
stime