6-2
Return to
2600AS-901-01 Rev. B / September 2008
Section 6: Range, Digits, Speed, Rel, and Filters
Series 2600A System SourceMeter® Instruments Reference Manual
Overview
The documentation in this section provides detailed information on characteristics and script
programming for each of the following functions:
•
•
•
•
•
Range
The selected measurement range affects the accuracy of the measurements as well as the
maximum signal that can be measured. Note that dashed lines are displayed (for example, --.----
µA), to indicate that the previous measurement is not recent. This usually happens when a change
occurs such as selecting a different range.
Available ranges
lists the available source and measurement ranges for the Keithley Instruments Series
2600A System SourceMeter
®
instruments.
Table 6-1
Source and measurement ranges
Model 2601A/2602A
Model 2611A/2612A
Model 2635A/2636A
Voltage Ranges Current Ranges
Voltage Ranges Current Ranges Voltage Ranges
Current Ranges
100mV
1V
6V
40V
100nA
1
μ
A
10
μΑ
100
μΑ
1mA
10mA
100mA
1A
3A
200mV
2V
20V
200V
100nA
1
μ
A
10
μΑ
100
μΑ
1mA
10mA
100mA
1A
1.5A
10A
1
200mV
2V
20V
200V
100pA
2
1nA
10nA
100nA
1
μ
A
10
μΑ
100
μΑ
1mA
10mA
100mA
1A
1.5A
1. 10A range available only in pulse mode.
2. 100pA range only in measure.