Series 2600A System SourceMeter® Instruments Reference Manual
Section 20: Calibration
src_FS
-- FS 2600A source output value
DMM_FS_rdg
-- FS DMM measurement
Typical values for the Model 2601A/2602A 1V range:
smua.source.calibratev(-1,-1e-10,-1e-4,-0.9,-0.896)
Typical values for the Model 2611A/2612A/2635A/2636A 2V range:
smua.source.calibratev(-2,-1e-10,-1e-4,-1.8,-1.805)
22) If this step is not on the CALA sense mode, send the measure calibration command
using the multimeter and Series 2600A readings, and range setting for the
parameters:
smua.measure.calibratev(-range,Z_rdg,DMM_Z_rdg,
FS_rdg,DMM_FS_rdg)
Where:
range
-- negative of the present calibration range
Z_rdg
-- zero 2600A measurement
DMM_Z_rdg
-- zero DMM measurement
FS_rdg
-- FS 2600A measurement
DMM_FS_rdg -- FS DMM measurement
Typical Model 2601A/2602A 1V range values:
smua.measure.calibratev(-1,-1e-4,-1e-6,-0.89,-0.896)
Typical Model 2611A/2612A/2635A/2636A 2V range values:
smua.measure.calibratev(-2,-1e-4,-1e-6,-1.81,-1.805)
d.
Be sure to complete steps a through v for all six voltage steps in
before continuing with current calibration.
e.
Select auto polarity mode:
smua.cal.polarity = smua.CAL_AUTO