2600AS-901-01 Rev. B / September 2008
Return to
19-255
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
Output
Data
Pulsed current measurements are stored in the reading buffer specified by the
buffer
input argument.
Example
ConfigPulseVMeasureI(smub,
0,
20,
1,
0.001,
0.080,
10,
smub.nvbuffer1, 2)
Set up a pulse train that will use SourceMeter instrument channel B. The pulse
amplitude will be 20V and will return to 0V after 1ms. The pulse will remain at 0V
for 80ms and the current limit will be 1A during the pulse. The pulse train will
consist of 10 pulses and the pulse train will be assigned a tag index of 2.
See also
,
Table 19-14
KIPulse TSP test script: ConfigPulseIMeasureVSweepLin
TSP project name:
KIPulse
TSP test script name:
KIPulse
Function:
ConfigPulseIMeasureVSweepLin
Usage:
f, msg = ConfigPulseIMeasureVSweepLin
(smu,bias,start,stop,limit,ton,toff,points,buffer,tag[,sync_in]
[,sync_out][,sync_in_timeout][,sync_in_abort])
Description
Configures a linear pulsed current sweep with a voltage measurement at each
point. Measurement(s) will be made at the end of the
ton
time. The magnitude
of the first pulse will be
start
amps. The magnitude of the last pulse will be
stop
amps. The magnitude of each pulse in between will be
step
amps larger than the
previous pulse where:
step = (stop - start) / (points - 1)
This function does not cause the specified
smu
to output a pulse train. It simply
checks to see if all of the pulse dimensions are achievable and if so, assigns the
indicated
tag
or index to it. The
InitPulseTest(tag)
and
InitPulseTestDual(tag)
functions are used to initiate a pulse train assigned
to a valid
tag
.
Table 19-13 (cont.)
KIPulse TSP test script: ConfigPulseVMeasureI
TSP project name:
KIPulse
TSP test script name:
KIPulse
Function:
ConfigPulseVMeasureI
Usage:
f, msg = ConfigPulseVMeasureI
(smu,bias,level,limit,ton,toff,points,buffer,tag[,sync_in][,sync_out]
[,sync_in_timeout][,sync_in_abort])