1-12
Return to
2600AS-901-01 Rev. B / September 2008
Section 1: Getting Started
Series 2600A System SourceMeter® Instruments Reference Manual
1. CHANNEL A and CHANNEL B (Channel B on Model 2636A only)
Triax connectors for input/output, guard, and sense connections. Use only low-noise triax cables such as
the Keithley Instruments Model 7078-TRX (available in several lengths). Connector terminals and
associated triax cable conductors are as follows:
WARNING
When connecting to the model 2611A, 2612A, 2635A and 2636A
SMU outputs, with cables not rated for voltages above 42V, such as
the 2600A-ALG-2, you must disable the high voltage output by
using the INTERLOCK function as defined in section 10 of this
manual. Leaving the high voltage enabled while not properly
insulating the external connections to the unit poses a shock
hazard which could cause serious injury to the user. It is also
recommended that the LO connection terminal not be allowed to
float by connecting it to signal ground or another known signal
reference.
2. DIGITAL I/O
Female DB-25 connector. Fourteen pins for digital input or output, one pin for safety interlock. Use a cable
equipped with a male DB-25 connector (Keithley Instruments part number CA-126-1CA).
3. IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the Model 7007-1 or Model
7007-2.
4. Cooling exhaust vent
Exhaust vent for internal cooling fan. Keep vent free of obstructions to prevent overheating.
5. Chassis ground
Ground screw for connections to chassis ground.
6. RS-232
Female DB-9 connector. For RS-232 operation, use a straight-through (not null modem) DB-9 shielded
cable for connection to the PC (Keithley Instruments Model 7009-5).
7. TSP-Link
Expansion interface that allows a Series 2600A and other TSP-enabled instruments to trigger and
communicate with each other. Use a category 5e or higher LAN crossover cable (Keithley Instruments
part number CA-180-3A).
8. Power module
Contains the AC line receptacle and power line fuse. The instrument can operate on line voltages of 100V
to 240V AC at line frequencies of 50Hz or 60Hz. See
of this manual for line fuse replacement
instructions.
Table 1-1
Connectors and triax cable conductors
Connector
Center conductor
Inner ring
Outer ring
LO
Sense LO
Input/Output LO
Chassis ground
HI
Input/Output HI
Guard
Chassis ground
SENSE HI
Sense HI
Guard
Chassis ground
Triax cable
Center conductor
Inner shield
Outer shield