19-196
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
status.request_event
Attribute
Service request event register.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
Reads the service request event register:
serveventreg = status.request_event
Remarks
• These attributes are used to read the service request event register, which is returned as a
numeric value. Reading this register returns a value. The binary equivalent of the returned
value. The least significant bit of the binary number is bit 0, and the most significant bit is bit 7.
• For example, assume value 129 is returned for the event register. The binary equivalent is
10000001. This value indicates that bit B0 (MSB) and bit B7 (OSB) are set.
• The service request event register uses most of the same summary events as the status byte.
Bit B6 (MSS) is not used by the event register. For details, see
.
Details
Status byte and service request (SRQ)
Example
Reads the service request event register:
serveventreg = status.request_event
print(serveventreg)
Output:
1.02
The above output indicates that bits B0 (MSS) and B7 (OSB) are set.
status.reset
Function
Resets all bits set in the status model.
Usage
status.reset()
Remarks
This function clears all status data structure registers (enable, event, NTR, and PTR) to their
power up states.
Details