Table 4
Guidance and manufacturer’s declaration – electromagnetic immunity –
for equipment and systems that are not life-supporting
The ENDOFLATOR
®
40 model UI400 is intended for use in the electromagnetic environment speci
Ƃ
ed below.
The customer or user of the ENDOFLATOR
®
40 model UI400 should ensure that it is used in such an environment.
Immunity test
IEC 60601 test
|
level
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the ENDOFLATOR
®
40 model UI400, includ-
ing cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
Conducted RF
disturbances
IEC 61000-4-6
3 V
rms
150 kHz to 80 MHz
3
V
rms
B
Radiated RF disturbances
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3
V/m
B
80 MHz to 800 MHz
E
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
Field strengths from
Ƃ
xed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked with
the following symbol:
7
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re
ƃ
ection by structures, objects
and people.
a
Field strengths from
Ƃ
xed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
Ƃ
xed RF
transmitters, an electromagnetic site survey should be considered. If the measured
Ƃ
eld strength in the location in which the unit is used exceeds
the applicable RF compliance level above, the unit should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the unit.
b
Over the frequency range 150 kHz to 80 MHz,
Ƃ
eld strengths should be less than
3
V/m.